Product Description
The Alcatel ASM 180 TD is a high-performance helium leak detector designed for industrial applications requiring precise and reliable leak detection . This “dry” oil-free model makes it ideal for cleanroom environments and applications where hydrocarbon contamination must be avoided .
The detector utilizes a robust vacuum circuit incorporating a primary roughing pump, a molecular drag pump, a hybrid turbomolecular pump, an analyzer cell, and a calibrated leak module . It operates in three stages for vacuum test mode—primary roughing, molecular roughing, and fine leak test mode—and also supports sniffing mode (LDS) and internal calibration mode .
The autocalibration system uses an internal calibrated leak (approximately 1×10⁻⁷ mbar.l/s) equipped with a temperature-dependent compensation system, ensuring measurement accuracy and reliability . The calibration process is fully automated by pressing the autocal button .
Key Specifications
| Specification | Details |
|---|---|
| Manufacturer | Alcatel (now Pfeiffer Vacuum) |
| Model | ASM 180 TD |
| Type | Helium Mass Spectrometer Leak Detector |
| Pumping System | Dry (oil-free) primary roughing pump with molecular drag and turbomolecular pumps |
| Sensitivity | Very high sensitivity (2×10⁻¹¹ mbar.l/s) |
| Helium Pumping Speed | 4.4 l/s at inlet port |
| Autocalibration | Yes – with integrated calibrated leak |
| Roughing Pump Options | 4 m³/h (2.4 cfm) standard |
| Test Modes | Vacuum test (primary roughing, molecular roughing, fine leak), Sniffing mode (LDS) |
| Interface | RS-232, optional USB/Ethernet |
| Optional Features | Alphanumeric Control and Display Panel (ACDP), remote control unit, stainless steel cover, “l” gas line test option, 50 m³/h roughing |
| Power Consumption | 1.2 kVA |
| Voltage | 200-220-240 V, 50/60 Hz |
| Noise Level | 54 dB |
| Dimensions | 16.9″ x 17.9″ x 23.6″ |
Applications
-
Semiconductor Manufacturing: Leak testing of process tools and vacuum chambers
-
Aerospace: Testing components, pressure vessels, and sealed systems
-
Pharmaceutical: Ensuring integrity of packaging and processing equipment
-
Automotive: Leak testing of HVAC systems, fuel lines, and components
-
Research & Development: Precision leak measurement in laboratory settings
-
Industrial Maintenance: Detecting leaks in piping, vessels, and compressed air systems
-
Gas Line Testing: Optional “l” gas line test for long lines with reduced response time